SOTOX A™ MOBILE ANALY ZER | USER GUIDE
35 | EN
11. SHUTDOWN PROCEDURE | 12. TECHNIC AL SUPPORT, CODES AND TROUBLESHOOTING
11.
Shutdown Procedure
11.1
Turning the analyzer off
Step 1: Remove any SoToxa™ Test Cartridges from the SoToxa™ Mobile Analyzer.
Step 2: Press the power button on the right hand side of the analyzer.
Step 3: The screen will read ‘Analyzer Shutting Down’.
Step 4: When the screen turns dark, the analyzer has switched off.
The analyzer cannot be switched off if a test is in progress or if a test cartridge is inserted.
11.2
Automatic shutdown
In order to save power, the SoToxa™ Mobile Analyzer can be set to automatically shutdown after a period of inactivity. Refer to section 10.4.6
Automatic shutdown
.
The analyzer will not automatically shutdown if a test is in progress or if a test cartridge is inserted.
12. Technical Support, Codes and Troubleshooting
12.1
Technical support
In the event of an incomplete test or SoToxa™ Mobile Analyzer issue, in most cases on screen feedback will be provided to the user to assist
with diagnosis and resolution of the issue. In some cases, the analyzer will report specific codes – please refer to the list of codes on the
following pages for interpretation. If the issue cannot be resolved using this guide, please contact Customer Service.
Examples of screens:
Step 3