Pima™ ANALYSER
USER GUIDE
17
Inserting the Pima Test Cartridge
Insert cartridge into the Pima Analyser in the direction indicated by the arrow
on the cartridge.
The Pima Analyser is designed to be used in combination with the Pima
test cartridge only. No other cartridges, test strips or devices should be
inserted into the Pima Analyser.
Do not discard the cartridge pouch until the Pima Analyser has
successfully started the analysis. In case of a barcode error, the information
printed on the cartridge pouch can be used to enter the barcode manually
(refer to p. 40 for Barcode Error message).
•
Upon inserting the Pima test
cartridge a sensor in the Pima
Analyser recognizes that a
cartridge has been inserted and
automatically starts the analysis
process by fully drawing the
cartridge into the Pima Analyser
and closing the cartridge slot door.
•
The display screen changes briefly
to «Reading cartridge» at this
moment.
•
Once the Pima Analyser has accepted the cartridge it will prompt the
operator to enter both the Operator and Sample ID.
During data entry the test analysis continues in the background.
While it is not necessary to enter Operator and Sample ID before starting
the analysis, the Pima Analyser will NOT display a result until both
Operator and Sample ID have been entered.
2020-11-01
11:05
ANALYSIS
Reading
cartridge