Pima™ ANALYSER
USER GUIDE
23
Abort Analysis
•
Select «Abort Analysis» in the «Analysis Menu» and confirm with
〈
.
Note that cartridges CANNOT be re-used after an analysis is aborted.
•
The «Abort Analysis» window
appears and «NO» is selected by
default. Press either
〈
or
⌨
to exit this window and to return
to the «Analysis Menu».
•
Select «YES» and confirm with
〈
.
The Pima Analyser will abort the
current analysis.
•
The operator is prompted to
remove the cartridge. The following
error code is displayed in the first
result window.
Press
⌨
to return to
the initial «Run test» window and
to start the next analysis.
To abort the export, deletion or printing of results, proceed in the same
manner. Press
⌨
to open the «Abort process» window. «NO» is selected by
default. Select «YES» and confirm with
〈
.
The Pima Analyser will abort the current process.
2020-11-01
11:05
Do you want to
abort the
analysis ?
ABORT ANALYSIS
YES
NO
2020-11-01
11:05
Do you want to
abort the
analysis ?
ABORT ANALYSIS
YES
NO
11:05
2020-11-01
Sample: 67352
Manual abort
Error 206
_________________________________________
BACK X PRINT