14
Switching On the Pima Analyser/Power Status Light
The green power status light located on the front of the Pima Analyser
displays its power status.
Press and hold the Power On button until you hear a triple "beep" and the
appearance of «pima
TM
» on the screen confirms that the Pima Analyser is now
switched on.
An automated initialization and self-test procedure is activated which may last
for approximately two minutes. During this time the screen changes and the
function of the cartridge slot door is tested.
If the Pima Analyser fails correct initialization and self-test, an error code is
displayed. It may not be possible to turn the Pima Analyser off via the route
described in section 7. In such a case press and hold the Power On
button on the back of the Pima Analyser until the screen turns black and you
hear a single "beep", confirming that the Pima Analyser is now switched off.
Restart the Pima Analyser as described above. Refer to section 8 and 9 for
details on error codes and contact your local distributor or Technical
Support, if necessary.
It is recommended to shutdown the Pima Analyser at least once per month.
Pima Analyser On
On
Pima Analyser Off
Off
The charging status light will only be illuminated when connected to main
A/C power. It may take several minutes until the correct charging status is
displayed.
In order to ensure maximum battery life charge the analyzer battery fully
overnight (8 -10 hours) before running the
Pima Analyser
on battery power.
When running the
Pima Analyser
on battery power, the battery status is
displayed by the battery status symbol within the display status line.