9
1
2
3
4
5
6
7
8
Pima™ ANALYSER
USER GUIDE
39
Error Codes
Error Code
Action
200
Switch off the Pima Analyser via Main Menu and restart.
If Error persists contact your local distributor.
201
Check cartridge control window was fully filled.
Repeat test with new cartridge.
202
Check cartridge cap was completely closed.
Repeat test with new cartridge.
203
Check cartridge expiry date.
Check correct date on the Pima Analyser.
Repeat test with valid cartridge.
204
Repeat test with new cartridge.
205
Contact your local distributor.
206
Analysis aborted by operator. No further action required.
207
Repeat test with new cartridge.
If Error persists contact your local distributor.
210, 220, 230
Repeat test with new cartridge.
If Error persists contact your local distributor.
211
Switch off the Pima Analyser via Main Menu and restart.
If Error persists contact your local distributor.
300-399
Contact your local distributor.
810
820
825
830
840
850
860
Repeat test with new cartridge.
If Error persists contact your local distributor.
870
880
910
920
930
940