m-PIMA™ ANALYSER
USER GUIDE
63
9
Error Code
Action
21111-21115, 21120
Make sure that the sample capillary is
completely filled. Keep analyser door closed
during test run.
Repeat test with new cartridge.
Contact your local distributor or your regional
Technical Support if the error persists.
21501
Repeat test with new cartridge.
Make sure that the sample capillary is
completely filled.
21502
Too low atmospheric pressure. Do not continue
with testing until you have reached altitudes that
are within specifications.
21503
Too high atmospheric pressure. Do not continue
with testing until you have reached altitudes that
are within specifications.
23020
Cartridge or sample related errors.
Please refer to the test cartridge guides PI-
m-PIMA-04 or PI-m-PIMA-05 for further
information about sample storage and
handling.
23028
Unexpected amplification curve, e.g. due to
individual sample properties.
Repeat test with new cartridge.
Contact your local distributor or your regional
Technical Support if the error persists.