Table 9
displays the concurrence between the GM and YSI reference glucose when GM reads ‘HI’. For
example, when GM reads ‘HI’ you can expect that YSI glucose values were less than 200 mg/dL 0.0%
of the time, between 200 and 300 mg/dL 0.0% of the time, between 301 and 400 mg/dL 0.0% of the
time, between 401 and 500 mg/dL 100% of the time, and above 500 mg/dL 0.0% of the time.
Table 9: Concurrence Analysis with HI GM Reading
GM Glucose Level
(mg/dL)
YSI (mg/dL)
N
<200
200-300
301-400
401-500
>500
>500 (HI)
0.0
0.0
0.0
100.0
0.0
2
57