Contraindications
The FreeStyle Libre Pro Flash Glucose Monitoring System
must be removed prior to Magnetic Resonance Imaging (MRI),
Computed Tomography (CT) scan, or high-frequency electrical
heat (diathermy) treatment . The effect of MRI, CT scans, or
diathermy on the performance of the System has not been
evaluated . The exposure may damage the Sensor and may
impact proper function of the device to detect trends and track
patterns in the user’s glucose values during the wear period .
WARNING:
The FreeStyle Libre Pro Flash Glucose Monitoring
System contains small parts that may be dangerous if
swallowed .
CAUTION:
•
Performance of the System when used with other implanted
medical devices, such as pacemakers, has not been evaluated .
•
Some individuals may be sensitive to the adhesive that
keeps the Sensor attached to the skin . If your patient notices
significant skin irritation around or under their Sensor, they
should remove the Sensor and stop using the FreeStyle Libre
Pro System . Follow your facility’s procedures for handling
skin reactions .