THERMAL MASS FLOWMETER | OI/FMT230/250-EN REV. C
… 8 Commissioning and operation
… Interface description
Parameter range – Diagnostics
Modbus register
address
Parameter name
Data type [register length] / value
range
Description
... / ...Diagnosis Control
3313
Preset Maint. cycle
TUSIGN32 [2]
0 to 99999 h
Sets the service interval.
After the maintenance interval has expired, the corresponding error
message ‘Maintenance interval is reached’ is set. The setting ‘0’
deactivates the maintenance interval.
Factory setting: 0 h
3311
Maint. Remain. Time
TUSIGN32 [2]
Time remaining in the maintenance interval until the error message
‘Maintenance interval is reached’ is set.
The parameter is read only.
9001
Start New Cycle
ACTION [1]
Resetting of the maintenance interval.
By writing any value to this address, the maintenance interval is reset to
the value set under ‘Preset Maint. cycle’.
... / ...Diagnosis Values
247 Gas
Temperature
TFLOAT [2]
Output of current meter tube temperature in °C.
The parameter is read only.
223 Transmitter
Temp.FE
TFLOAT [2]
Output of current housing temperature in °C.
The parameter is read only.
bis / ...Simulation Mode
4001 Simulation
Switch
TUSIGN8 [1]
0: Off
1:
Qm mass [unit]
2: Temperature
[unit]
3:
Qv @Vol.flow [Unit]
4: Density@
[unit]
50: Qm mass [%]
51: Temperature
[%]
52: Qv @Vol.flow [%]
120: Digital output 41/42
121: Digital output 51/52
Manual stimulation of measured values / outputs.
The simulated output values correspond to the set measured value (see
Setting of the simulated measured values
on page 87).
Only one measured value / output can be selected for simulation.
After power-up / restart of the device, the simulation is switched off.