Sample Gas Inlet Conditions
Sample Gas Inlet Conditions
Sample gas pressure
p
abs
= 70 to 105 kPa (0.7 to 1.05 bar)
Sample gas flow rate
Gas path: 7 to 70 l/h
Bypass: 15 to 150 l/h (not in versions with Peltier cooler)
Sample gas inlet temperature +10 to +50 °C
Sample gas inlet dew point
Version without Peltier cooler: The sample gas dew point should be at least 5 °C
below the lowest ambient temperature throughout the sample gas path. If this is
not the case, then a sample gas cooler unit or a condensate trap is required.
Version with Peltier Cooler: max. 50 °C (at max. +25 °C ambient temperature)
Scope of Delivery
Scope of Delivery
Quantity Description
1
SCC-S Sample gas feed unit
1
Power supply cable, length 2 m
1
Socket plug, 8-pole, for signal line connection
2
Operator’s manuals (1 copy in German, 1 in English)