22
Replacing the Condensate Pump Pressure Rollers and Springs
When do the Pressure Rollers and Springs need to be Replaced?
The pressure rollers in the condensate pump must be replaced when their surface
is damaged.
The pressure springs in the condensate pump must be replaced when they are
broken.
NOTE
All four pressure springs must be replaced at the same time to ensure trouble-free
operation.
WARNING
Residues from the condensate that the pump has been feeding may be found
in the pump hose and the connected hoses. These media can flow out when
the hose connections are opened. Take appropriate measures where needed to
collect such residues. The medium being fed may be corrosive and poisonous!
Appropriate precautions must be taken! Wear protective goggles and
protective clothing!
Condensate Pump, Roller Mounting
1
Moving belt
5
Roller mounting
2
S-clip
6
Pressure springs (x 4)
3
Screws for securing the pump head (x 2)
7
Roller axle
4
Pump head
8
Pressure roller (x 2)