Shutting Down the Sample Gas Feed Unit
WARNING
Before the sample gas feed unit is shut down it must be purged to prevent the
accumulation of condensate and deposits.
NOTE
Make sure the sample gas feed unit is free of residual moisture that can freeze if
low temperatures are encountered during storage and transport. Ambient
temperature during storage and transport: –25 to +60 °C.
Shut Down the Sample Gas Feed Unit
1
Disconnect the power supply.
2
Shut off the sample gas supply to the sample gas feed unit.
3
Remove the gas pipes from the connections on the feed unit.
4
Thoroughly purge the sample gas feed unit gas paths with an inert gas.
5
Fully tighten the gas connections.
6
Disconnect the electrical leads.