Messages and tests
40 - EN
FSM4000
D184B141U02
Error code and
clear text message
Prior-
ity
Description
Possible cause
Corrective action
Error: A
Max. Alarm
10, 11 Max alarm limit value. The configured Max alarm
for the flowrate was
exceeded.
Reduce flowrate.
Error: B
Min. Alarm
Min alarm limit value. The flowrate is below the
configured Min alarm
Increase flowrate.
Error: C
Ext. data base
3 External
database
FRAM is defective or
not available.
Missing FRAM or
defective FRAM
1)
.
Install and screw in place the
FRAM for the associated
flowmeter sensor to the
connection board in the field
housing unit. See chapter 5. If
the FRAM cannot be read, it
must be replaced.
Error: D
Old Primary
12 Flowmeter
sensor
type
from the older model
flowmeter series was
selected.
Configuration under the
parameter "Primary
Setup" is incomplete.
Complete configuration under
"Primary Setup". See also
chapter 11 at the operating
instruction.
Error: E
DC to high
13 Increased
analog
reset, measurement
signal with large DC
Air bubbles, deposits on
electrodes, too high
interference signal.
Empty pipe.
Activate "Noise reduction" in
the operating mode submenu.
Use a air separator, clean
electrodes, activate empty pipe
detector.
Contact ABB Service.
Error: F
FRAM in Primary
1
Missing data from
external FRAM.
FRAM for the flowmeter
sensor is still installed in
the terminal box.
Install and screw in place the
FRAM for the associated
flowmeter sensor to the
connection board in the field
housing unit. See chapter 5.
Min-Max over/undershoot
for diagnostic values.
Retrieve details of error
message under the Status
submenu in the Diagnostic error
register. Adjust thresholds, if
necessary.
Error: G
Error Diagnosis
14
Diagnostic or wiring
error
Wiring error for electrode
circuit. Coil short-circuit
to ground
Retrieve details of error
message under the Status
submenu in the Wiring error
register.
1)
Replacing a defective FRAM.
If the FRAM is defective and startup cannot be performed, a new FRAM can be requested from the Göttingen
plant. To properly process requests, you must include the ABB order number and unit number of the flowmeter
sensor. After installing the FRAM and switching on the power, you can start up the system. Review all
flowmeter sensor data and setup data for the system, and enter this information again, if necessary.