Parameterization
D184B141U02
FSM4000
EN - 35
6 Parameterization
6.1 Data
entry
Use the keys (4) to enter data when housing is open. If closed, use the magnet stick (5) and the
magnet sensors. The stick is held over the appropriate NS symbol.
G00566
N
S
N
S
N
S
FSM4000
DATA
STEP
C/CE
ENTER
1
2
3
4
5
Fig. 39
1
Magnet sensor C/CE
2
Magnet sensor DATA/ENTER
3
Magnet sensor STEP
4
Control buttons
5
Magnet
When entering data, the transmitter remains online, i.e., current and pulse outputs show the
current operating mode. The functions of the individual keys are explained in the following:
C/CE
Toggle between operating mode and menu.
STEP
↓
The STEP key is one of two arrow keys. Use STEP to scroll forward
through the menu. All the desired parameters can be called up.
DATA
↑
The DATA key is one of two arrow keys. Use DATA to scroll backward
through the menu. All the desired parameters can be called up.
ENTER The ENTER function requires that both arrow keys, STEP and DATA, be
pressed simultaneously. ENTER has the following functions:
• Access the parameter to be changed and set the new, selected or
default parameter.
The ENTER function is effective for approx. 10 s only. If a new value is not
entered within 10 s, the display reverts to the old value.