Electrical
connection
D184B141U02
FSM4000
EN - 27
Pulse output
Pulse output active
Pulse output, passive, optocoupler
R
B
*
52
G00247
51
R
B
*
U
CE
I
CE
+ 16 ... 30 V DC
-
I
E
Function:
fmax:
fmin:
Setting range:
active/passive, selectable via jumper (see "Commissioning" in operating instructions)
5 kHz,
0.00016 Hz
Pulses/unit, pulse width (observe dynamic limits)
Active:
150
Ω
≤
Load < 10 k
Ω
Pulse
width
≤
50 ms, max. pulse frequency
≤
3 Hz,
500
Ω
≤
Load < 10 k
Ω
Pulse
width
≥
0.1 ms, max. pulse frequency: 5 kHz
Passive:
"closed": 0 V
≤
UC
EL
≤
2 V, 2 mA
≤
I
CEL
≤
220 mA
"open": 16 V
≤
U
CEH
≤
30 V, 0 mA
≤
I
CEH
≤
2 mA
Terminals: 51,
52
Fig. 32: Pulse output, active and passive, optocoupler
PROFIBUS PA / FOUNDATION Fieldbus
G00248
2
FF-
FF+
R
C
1
97
PA-
R
C
PA+
98
97
98
I
E
I
E
The resistance R and condenser C form the bus termination. They
must be installed when the device is connected to the
end of the entire bus cable.
R = 100
Ω
; C = 1 µF
1
PROFIBUS PA
2
FOUNDATION Fieldbus
I = Internal
E = external
Fig. 33
Bus termination for S4 transmitter
To terminate the bus if the instrument is at the end of the bus cable, the termination components in the S4 transmitter can be used. To do so,
close both hook switches in the connection area of the transmitter.
Important
If the transmitter module is removed, bus termination is also canceled.
G00249
M3
M1
41
42
97
98
U+
2S
1
1S
U-
3
3
2
16
6
7
Connection board for field housing unit
Hook switches for
bus termination must be closed
Fig. 34