COM/AZ40/MOD–EN Rev. A

| Endura AZ40 | Oxygen and carbon monoxide equivalent (COe) analyzer

A.3.5 Sensor signals view – COe signals view page 3

A.3.6 Sensor signals view – raw signals view page 4

A.3.7 Calibration timer

A.3.8 Analog outputs

A.4 Read only 32 bits

A.4.1 Transmitter date

Number of seconds since 01/01/2000

A.4.2 Scheduled events

number of seconds since 01/01/2000

A.5 Read only 16 bits

A.5.1 Sensor hardware version

Valid range 1 to 9.

A.5.2 Mains voltage

A.6 Read-only 8 bits

A.6.1 Sensor type

A.6.2 Media card version

A.7 Read-only strings

MODBUS address

Description

30068

View 1 – COe concentration

30070

View 2 – COe element temperature difference (°)

30072

View 3 – COe active element temperature (°)

30074

View 4 – COe reference element temperature (°)

30076

View 5 – COe sensor last zero

30078

View 6 – COe sensor last span

30080

View 7 – COe sensor active zero

30082

View 8 – COe sensor active span

MODBUS address

Description

30084

View 1 – Oxygen sensor output (mV)

30086

View 2 – Flange block (mV)

30088

View 3 – COe block (mV)

30090

View 4 – Inlet temperature

30092

View 5 – Outlet temperature

30094

View 6 – Transmitter temperature (°)

30096

View 7 – Sensor cold junction temperature (°)

30098

View 8 – Electrical cold junction temperature (°)

MODBUS address

Description

30100

Calibration progress (% completion)

MODBUS address

Description

30102

Analog output value (mA) AOUT1

30104

Analog output value (mA) AOUT2

30106

Analog output value (mA) AOUT3

30108

Analog output value (mA) AOUT4

30110

Analog output % of full scale AOUT1

30112

Analog output % of full scale AOUT2

30114

Analog output % of full scale AOUT3

30116

Analog output % of full scale AOUT4

MODBUS address

Description

31000

Transmitter – Date of manufacture

MODBUS address

Description

31002

Sensor – Date of manufacture

31004

Scheduled calibration start time

31006

Scheduled blowback start time

31008

Next scheduled calibration time

31010

Next scheduled blowback time

MODBUS address

Description

32000

Sensor – Hardware version number

MODBUS address

Description

32001

Mains voltage selection

(0 = Not Set, 1 = 115V, 2 = 230V)

MODBUS address

Description

Definitions

33000

Blowback fitted

0 = Not Fitted

1 = Continuous

2 = Pulsed

33001

Sensor type

0 = Unrecognized

140 = Oxygen 

141 =  COe

255 = None

33002

Sensor calibration error

MODBUS address

Description

33003

Media card percentage full

33004

Media card hardware version number

MODBUS address

Description

No. of 

Bytes

34000

Transmitter hardware revision number

2

34002

Transmitter software revision number

19

34021

Transmitter serial number

14

34035

Sensor software revision number

19

34054

Sensor serial number

15

34069

Media unit software revision number

19

 

Summary of Contents for Endura AZ40

Page 1: ...s together with MODBUS reference tables for the Endura AZ40 analyzer For more information Further publications for the Endura AZ40 analyzer are available for free download from www abb com analytical...

Page 2: ...f specification alarms 10 A 1 6 Transmitter and sensor diagnostic maintenance alarms 10 A 1 7 Media card diagnostics 11 A 1 8 Relay state diagnostics 11 A 1 9 Alarm state diagnostics 11 A 2 Digital ou...

Page 3: ...re injury WARNING Bodily injury High temperature This symbol in conjunction with the signal word WARNING indicates a potentially dangerous situation Failure to observe this safety information may resu...

Page 4: ...tested half mask conforming to EN140 or equivalent with a level 3 particulate filter conforming to EN 143 or equivalent Disposable protective coveralls in accordance with Type 5 ISO 13982 1 2004 or e...

Page 5: ...btained from the Company together with servicing and spares information WARNING Bodily injury The transmitter weighs 7 6 kg 17 lb and must be mounted in accordance with the information supplied in the...

Page 6: ...c disposal systems after 12 August 2005 To conform to European local and national regulations EU Directive 2002 96 EC European electrical equipment users must now return old or end of life equipment t...

Page 7: ...below the Select prompt The Communication page appropriate to the communication module fitted is displayed MODBUS refer to Section 4 page 8 IMPORTANT NOTE Communication level menus are enabled only if...

Page 8: ...transmitter on a MODBUS link RS485 Setup Mode Select the MODBUS serial communication serial link type 2 Wire 4 Wire or Off 4 Wire Baud Rate Select a communication transfer rate 1200 2400 4800 9600 19...

Page 9: ...tions Make connections to the MODBUS serial board as shown in Fig 4 1 On systems with multiple slaves make connections to two or four wire link configurations in parallel as shown in Fig 4 2 When conn...

Page 10: ...ocouple failed 6 28 Sensor non volatile memory failed 7 29 Processor board non volatile memory failed 8 30 Main board non volatile memory failed 9 31 Reserved 10 32 Reserved 11 33 Comms board non vola...

Page 11: ...15 Acknowledge alarm 8 MODBUS address Description 30000 Oxygen concentration value 30002 COe concentration value 30004 Inlet temperature value 30006 Outlet temperature value 30008 Outlet Inlet tempera...

Page 12: ...rature MODBUS address Description 30100 Calibration progress completion MODBUS address Description 30102 Analog output value mA AOUT1 30104 Analog output value mA AOUT2 30106 Analog output value mA AO...

Page 13: ...w 40048 Analog output 3 Electrical range high 40050 Analog output 3 Electrical range low 40052 Analog output 3 Engineering range high 40054 Analog output 3 Engineering range low 40056 Analog output 4...

Page 14: ...couple types 43006 Efficiency calculation 0 Disabled 1 Enabled 43007 Blowback interval 0 Off 1 24 hours 43008 Reserved 43009 Scheduled calibration frequency units 1 Off 1 Daily 2 Weekly 3 Monthly 4301...

Page 15: ...ht saving 1 First 5 Last 2 Second 3 Third 4 Fourth 43034 Daylight saving end day for daylight saving 1 to 7 where 1 Sunday and 7 Saturday 43035 Daylight saving end month for daylight saving 1 to 12 wh...

Page 16: ...rrent enable 0 Disabled 1 Enabled 43061 Analog output 4 Source 0 None 6 Efficiency 43062 Analog output 4 Polarity 0 Inverted 1 Non Inverted 43063 Analog Output 4 Failure current enable 0 Disabled 1 En...

Page 17: ...ag Alarm 1 44096 Alarm 2 tag Alarm 2 44112 Alarm 3 tag Alarm 3 44128 Alarm 4 tag Alarm 4 44144 Alarm 5 tag Alarm 5 44160 Alarm 6 tag Alarm 6 44176 Alarm 7 tag Alarm 7 44192 Alarm 8 tag Alarm 8 MODBUS...

Page 18: ...18 COM AZ40 MOD EN Rev A Endura AZ40 Oxygen and carbon monoxide equivalent COe analyzer Appendix B Acknowledgements MODBUS is a registered trademark of the Modbus IDA organization...

Page 19: ...Endura AZ40 Oxygen and carbon monoxide equivalent COe analyzer COM AZ40 MOD EN Rev A 19 Notes...

Page 20: ...document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior writt...

Reviews: