| Endura AZ40 | Oxygen and carbon monoxide equivalent (COe) analyzer

Appendix A – MODBUS tables

A.1 Digital inputs

A.1.1 Sensor reading diagnostics

A.1.2 Transmitter function diagnostics

A.1.3 Transmitter and sensor diagnostic failure alarms

A.1.4 Transmitter and sensor diagnostic function alarms

A.1.5 Transmitter and sensor diagnostic 
out of specification alarms

A.1.6 Transmitter and sensor diagnostic maintenance alarms

MODBUS coil

Description

0

Oxygen concentration value status

1

COe concentration value status

2

Inlet temperature value status

3

Outlet temperature value status

4

Outlet / Inlet temperature value status

5

Efficiency value status

MODBUS coil

Description

6

Transmitter failure diagnostic

7

Transmitter out of specification diagnostic

8

Transmitter maintenance diagnostic

9

Transmitter function check diagnostic

10

Sensor failure diagnostic

11

Sensor out of specification diagnostic

12

Sensor maintenance diagnostic

13

Sensor function check diagnostic

14

Active failure alarm

15

Calibration in progress

16

Blowback in progress

17

Zero gas valve check

18

Span gas valve check

19

Test gas valve check mode

20

Sample flow rate test mode

MODBUS 

coil

Description

Diagnostic 

alarm no.

21

ADC Failed

0

22

Sensor electronics cold junction failed

1

23

Oxygen heater failed

2

24

Flange block thermocouple failed

3

25

COe block thermocouple failed

4

26

Inlet thermocouple failed

5

27

Outlet thermocouple failed

6

28

Sensor non-volatile memory failed

7

29

Processor board non-volatile memory failed

8

30

Main board non-volatile memory failed

9

31

Reserved

10

32

Reserved

11

33

Comms board non-volatile memory failed

12

34

S/W key 1 non-volatile memory failed

13

35

Reserved

14

36

Reserved

15

37

Internal communication failure between 

transmitter and sensor

16

38

Over temperature trip

17

MODBUS 

coil

Description

Diagnostic 

alarm no.

39

Flange, COe and oxygen heaters warming up

18

40

Flange, COe and oxygen heaters stabilizing

19

41

Calibrating

20

42

Blowback in progress

21

43

Sample flow rate test in progress

22

44

Manual valve test in progress

23

45

Remote zero gas test in progress

24

46

Remote span gas test in progress

25

47

Recovery stage after calibration / blowback / 

manual valve test / flow rate test

26

48

In hold mode

27

49

In simulation mode

28

MODBUS 

coil

Description

Diagnostic 

alarm no.

50

Last sensor calibration failed

29

51

Process value out of range

30

52

Temperature value out of range

31

53

Flange block temperature is high

32

54

Flange block temperature is low

33

55

COe block temperature is high

34

56

COe block temperature is low

35

57

Oxygen heater current is high

36

58

Oxygen heater current is low

37

MODBUS 

coil

Description

Diagnostic 

alarm no.

59

Transmitter internal ambient temperature is high

38

60

Transmitter internal ambient temperature is low

39

61

Sensor CJC temperature is high

40

62

Sensor CJC temperature is low

41

63

Sensor ECJC temperature is high

42

64

Sensor ECJC temperature is low

43

65

Sensor power is turned off via SW1

44

66

Sensor mains selection is in the wrong position

45

Summary of Contents for Endura AZ40

Page 1: ...s together with MODBUS reference tables for the Endura AZ40 analyzer For more information Further publications for the Endura AZ40 analyzer are available for free download from www abb com analytical...

Page 2: ...f specification alarms 10 A 1 6 Transmitter and sensor diagnostic maintenance alarms 10 A 1 7 Media card diagnostics 11 A 1 8 Relay state diagnostics 11 A 1 9 Alarm state diagnostics 11 A 2 Digital ou...

Page 3: ...re injury WARNING Bodily injury High temperature This symbol in conjunction with the signal word WARNING indicates a potentially dangerous situation Failure to observe this safety information may resu...

Page 4: ...tested half mask conforming to EN140 or equivalent with a level 3 particulate filter conforming to EN 143 or equivalent Disposable protective coveralls in accordance with Type 5 ISO 13982 1 2004 or e...

Page 5: ...btained from the Company together with servicing and spares information WARNING Bodily injury The transmitter weighs 7 6 kg 17 lb and must be mounted in accordance with the information supplied in the...

Page 6: ...c disposal systems after 12 August 2005 To conform to European local and national regulations EU Directive 2002 96 EC European electrical equipment users must now return old or end of life equipment t...

Page 7: ...below the Select prompt The Communication page appropriate to the communication module fitted is displayed MODBUS refer to Section 4 page 8 IMPORTANT NOTE Communication level menus are enabled only if...

Page 8: ...transmitter on a MODBUS link RS485 Setup Mode Select the MODBUS serial communication serial link type 2 Wire 4 Wire or Off 4 Wire Baud Rate Select a communication transfer rate 1200 2400 4800 9600 19...

Page 9: ...tions Make connections to the MODBUS serial board as shown in Fig 4 1 On systems with multiple slaves make connections to two or four wire link configurations in parallel as shown in Fig 4 2 When conn...

Page 10: ...ocouple failed 6 28 Sensor non volatile memory failed 7 29 Processor board non volatile memory failed 8 30 Main board non volatile memory failed 9 31 Reserved 10 32 Reserved 11 33 Comms board non vola...

Page 11: ...15 Acknowledge alarm 8 MODBUS address Description 30000 Oxygen concentration value 30002 COe concentration value 30004 Inlet temperature value 30006 Outlet temperature value 30008 Outlet Inlet tempera...

Page 12: ...rature MODBUS address Description 30100 Calibration progress completion MODBUS address Description 30102 Analog output value mA AOUT1 30104 Analog output value mA AOUT2 30106 Analog output value mA AO...

Page 13: ...w 40048 Analog output 3 Electrical range high 40050 Analog output 3 Electrical range low 40052 Analog output 3 Engineering range high 40054 Analog output 3 Engineering range low 40056 Analog output 4...

Page 14: ...couple types 43006 Efficiency calculation 0 Disabled 1 Enabled 43007 Blowback interval 0 Off 1 24 hours 43008 Reserved 43009 Scheduled calibration frequency units 1 Off 1 Daily 2 Weekly 3 Monthly 4301...

Page 15: ...ht saving 1 First 5 Last 2 Second 3 Third 4 Fourth 43034 Daylight saving end day for daylight saving 1 to 7 where 1 Sunday and 7 Saturday 43035 Daylight saving end month for daylight saving 1 to 12 wh...

Page 16: ...rrent enable 0 Disabled 1 Enabled 43061 Analog output 4 Source 0 None 6 Efficiency 43062 Analog output 4 Polarity 0 Inverted 1 Non Inverted 43063 Analog Output 4 Failure current enable 0 Disabled 1 En...

Page 17: ...ag Alarm 1 44096 Alarm 2 tag Alarm 2 44112 Alarm 3 tag Alarm 3 44128 Alarm 4 tag Alarm 4 44144 Alarm 5 tag Alarm 5 44160 Alarm 6 tag Alarm 6 44176 Alarm 7 tag Alarm 7 44192 Alarm 8 tag Alarm 8 MODBUS...

Page 18: ...18 COM AZ40 MOD EN Rev A Endura AZ40 Oxygen and carbon monoxide equivalent COe analyzer Appendix B Acknowledgements MODBUS is a registered trademark of the Modbus IDA organization...

Page 19: ...Endura AZ40 Oxygen and carbon monoxide equivalent COe analyzer COM AZ40 MOD EN Rev A 19 Notes...

Page 20: ...document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior writt...

Reviews: