![ABB Endura AZ40 Operating Instructions Manual Download Page 16](http://html1.mh-extra.com/html/abb/endura-az40/endura-az40_operating-instructions-manual_3312206016.webp)
16
COM/AZ40/MOD–EN Rev. A
| Endura AZ40 | Oxygen and carbon monoxide equivalent (COe) analyzer
A.11.6 Analog output settings
A.11.7 Digital I/O (relay) settings
MODBUS address
Description
Valid range
43052
Analog output 1 – Source
0 – None
1 – Oxygen
43053
Analog output 1 – Polarity
0 – Inverted
1 – Non-Inverted
43054
Analog output 1 – Failure current enable
0 – Disabled
1 – Enabled
43055
Analog output 2 – Source
0 – None
2 – Combustibles
43056
Analog output 2 – Polarity
0 – Inverted
1 – Non-Inverted
43057
Analog Output 2 – Failure current enable
0 – Disabled
1 – Enabled
43058
Analog Output 3 – Source
0 – None
5 – Temperature
43059
Analog output 3 – Polarity
0 – Inverted
1 – Non-Inverted
43060
Analog output 3 – Failure current enable
0 – Disabled
1 – Enabled
43061
Analog output 4 – Source
0 – None
6 – Efficiency
43062
Analog output 4 – Polarity
0 – Inverted
1 – Non-Inverted
43063
Analog Output 4 – Failure current enable
0 – Disabled
1 – Enabled
MODBUS address
Description
Valid range
43064
DO 1: Oxygen – source
15 – Alarm 1/2 state
43065
DO 1: Oxygen – polarity
0 – Inverted
1 – Non-inverted
43066
DO 2: Combustibles – source
16 – Alarm 3/4 state
43067
DO 2: Combustibles – polarity
0 – Inverted
1 – Non-inverted
43068
DO 3: Temperature – source
17 – Alarm 5/6 state
43069
DO 3: Temperature – polarity
0 – Inverted
1 – Non-inverted
43070
DO 4: Efficiency – source
18 – Alarm 7/8 state
43071
DO 4: Efficiency – polarity
0 – Inverted
1 – Non-inverted
43072
DO 5: Fault – source
28 – Analyzer fault
43073
DO 5: Fault – polarity
0 – Inverted
1 – Non-inverted
43074
DO 6: Calibration – source
27 – Calibration in progress
43075
DO 6: Calibration – polarity
0 – Inverted
1 – Non-inverted