Operating the Gas Analyzer
101
Key Functions in Measuring Mode
Switch over the display of each individual measured value; in
addition to the digital display, an analog bar with information on
the range limits is shown in this display.
Reduce or increase the contrast of the display.
When a status message is active, press key first.
OK
Switch to menu mode (see page 102).
If a status message is active : press the key to display the
message list (see page 122).
Number of Decimal Places
When the screen displays the measured value in physical units (e.g.
ppm) the number of places after the decimal point depends on the size of
the set measuring range:
Span
Places after the decimal point
0.05
5
0.5
4
5
3
50
2
500
1
> 500
0
For the display of the measured value as a percentage of the whole
measuring range (%Span) two places are always shown after the decimal
point.
When setting the parameters, the number of decimal places is the same
as in the display in measuring mode.