Preparation for Installation
21
Magnos206
Sample Gas Inlet Conditions
Tempe5…+50
°C
Positive pressure
2…100 hPa
Flow 30…90
l/h
Flammable Gases
The analyzer is suitable for the measurement of flammable gases in a
non-hazardous environment. The special requirements for the
measurement of flammable gases (see page 38) must be complied with.
Corrosive Gases
If the sample gas contains Cl
2
, HCl, HF or other corrosive components,
the analyzer may only be used if the sample gas composition has been
taken into account by the manufacturer for the configuration of the
analyzer.
Magnos27
Sample Gas Inlet Conditions
Tempe5…+50
°C
Positive pressure
2…100 hPa
Flow 20…90
l/h
Flammable Gases
The analyzer may not be used for the measurement of flammable gases.
Caldos27
Sample Gas Inlet Conditions
Tempe5…+50
°C
Positive pressure
2…100 hPa
Flow
10…90 l/h, min. 1 l/h
Pressure drop
< 2 hPa at 60 l/h N
2
Flammable Gases
The analyzer is suitable for the measurement of flammable gases in a
non-hazardous environment. The special requirements for the
measurement of flammable gases (see page 38) must be complied with.
Corrosive Gases
If the sample gas contains Cl
2
, HCl, HF, SO
2
, NH
3
, H
2
S or other corrosive
components, the analyzer may only be used if the sample gas
composition has been taken into account by the manufacturer for the
configuration of the analyzer.