CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
35
Process conditions
Temperature limits °C (°F)
Note
When using the device in potentially explosive atmospheres,
note the additional temperature data in
Temperature data
on
page 12!
Measuring medium temperature T
medium
• FCx430:
−
50 to 160 °C (
−
58 to 320 °F)
• FCx450:
−
50 to 205 °C (
−
58 to 401 °F)
In devices with order code ‘Extended tower length – TE3’, the
measuring medium temperature must be limited to a maximum
of 140 °C (284 °F) from an ambient temperature of
≥
65 °C
(149 °F).
Ambient temperature T
amb.
• Standard:
−
20 to 70 °C (
−
4 to 158 °F)
• Optional:
−
40 to 70 °C (
−
40 to 158 °F)
Pressure ratings
The maximum permissible operating pressure is determined by
the respective process connection, the temperature of the
medium to be measured, the screws, and the gasket material.
For an overview of available pressure ratings, see
Device designs
on page 26.
Housing as a protective device (optional)
Order code PR5
Maximum burst pressure 60 bar (870 psi)
Optional order code PR6 and PR7 on request
• Increased burst pressures up to 100 bar (1450 psi),
possible for nominal diameters DN 15 to 100 (½ to 4 in.).
• Increased burst pressures up to 150 bar (2175 psi),
possible for nominal diameters DN 15 to 80 (½ to 3 in.).
• Purge connections are available on request.
Pressure Equipment Directive
Conformity assessment in accordance with Category III, fluid
group 1, gas
Note the corrosion resistance of the meter tube materials in
relation to the measuring medium.
Material load for process connections
Note
You can reference the availability of the different process
connections in the Online ABB Product Selection Assistant (PSA)
for flow
www.abb.com/flow-selector
.
• Not all connections shown here are available in all the devices
and designs.
• The permissible material load of the device can additionally
differ from the material load of the connection. The
permissible limit values (pressure rating / measuring
medium temperature T
medium
) can be found on the name
plate.
Design Nominal
diameter
PS
max
TS
max
TS
min
Pipe fitting
(DIN 11851)
DN 15 to DN 40
(½ to 1½ in)
40 bar
(290 psi)
(580 psi)
140 °C
(284 °F)
−
40 °C
(
−
40 °F)
DN 50 to DN 100
(2 to 4 in)
25 bar
(290 psi)
(363 psi)
140 °C
(284 °F)
−
40 °C
(
−
40 °F)
Pipe fitting
(SMS 1145)
DN 25 to DN 80
(1 to 3 in)
6 bar (290
psi)
(87 psi)
140 °C
(284 °F)
−
40 °C
(
−
40 °F)
Tri-Clamp
(DIN 32676)
DN 15 to DN 50
(½ to 2 in)
16 bar (290
psi)
(232 psi)
120 °C
(248 °F)
−
40 °C
(
−
40 °F)
DN 65 to DN 100
(2½ to 4 in)
10 bar (290
psi)
(145 psi)
120 °C
(248 °F)
−
40 °C
(
−
40 °F)
ASME BPE Clamp
< DN 80
(< 3 in)
17.1 bar
(290 psi)
(248 psi)
121 °C
(249.8 °F)
−
40 °C
(
−
40 °F)
DN
80
(< 3 in)
15.5 bar
(290 psi)
(224.8 psi)
121 °C
(249.8 °F)
−
40 °C
(
−
40 °F)
DN
100
(< 4 in)
12.9 bar
(290 psi)
(187.1 psi)
121 °C
(249.8 °F)
−
40 °C
(
−
40 °F)