136
CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
… 10 Diagnosis / error messages
… Error messages
Error code / error text
Description
NAMUR classification
S049.019
Option Card 1 com error.
Communication error to plug-in card.
• Check that the plug-in card is installed correctly.
• Replace plug-in card if necessary.
• Contact ABB Service.
Out of specification
S048.020
Option Card 2 com error.
S047.0015
Pulse output is cut off.
The pulse rate or the frequency at the pulse output is outside the permissible limits.
• Check configuration of the parameters for the pulse output.
Out of specification
S046.000
Mass flowrate exceeds limits.
The mass flow is below or above the configured limit values ‘Qm Massflow Min’ and ‘Qm Massflow
Max’.
• Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and adjust if
necessary.
• Check
mass
flow.
Out of specification
S044.001
Volume flowrate exceeds limits.
The volume flow is below or above the configured limit values ‘Qv Volumeflow Min’ and ‘Qv
Volumeflow Max’.
• Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and adjust if
necessary.
• Check volume flow rate.
Out of specification
S043.036
Density exceeds min/max limits.
The density is below or above the configured limit values ‘Density Min’ and ‘Density Max’.
• Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and adjust if
necessary.
• Check
density.
Out of specification
S042.037
Medium temperat exceeds limits.
The measuring medium temperature is below or above the configured limit values ‘Temperature Min’
and ‘Temperature Max’.
• Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and adjust if
necessary.
• Check measuring medium temperature.
Out of specification
S041.039
Concentration inunit exceeds
The concentration in units is below or above the configured limit values ‘Concentrat. [u] Min’ and
‘Concentrat. [u] Max’.
• Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and adjust if
necessary.
• Check
concentration.
Out of specification
S040.040
Concentration in percent exceeds
The concentration in % is below or above the configured limit values ‘Concentrat. [%] Min’ and
‘Concentrat. [%] Max’.
• Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and adjust if
necessary.
• Check
concentration.
Out of specification