12 – EN
CORIOLISMASTER FCX100, FCX400
SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
Preparation
G12361
2x
2
3
3
2
1
1
5
4
Fig. 19
1. Loosen two of the four screws
5
on the foot and
replace with head hole screws
4
.
2. Unscrew both fixing screws
1
on the front and rear
cover of the dual-compartment housing and replace
with the holding rods
3
.
Lay seal wire
G12352
2
3
1
Fig. 20
1. Insert locking cylinder
3
on the holding rod
1
on the
rear housing cover.
2. Insert the seal wire
2
through the openings on the
holding rod
1
such that both ends are the same length.
3. Tightly twist the seal wire
2
right behind the locking
cylinder
3
to secure the locking cylinder
3
.
NOTICE
You should still be able to slightly move the locking
cylinder after it is secured with the seal wire.
4. Continue to twist the seal wire
2
to the holding rod
1
on the front housing cover. Both ends of the seal wire
must be very close together and the twist loops should
be smaller than 1 cm.
5. Secure the locking cylinder
3
on the front cover, as
described in step 3.
G12354
3
2
1
Fig. 21
6. Continue to twist the seal wire
1
up to the rear head
hole screw
3
. Guide one end of the wire through the
hole in the screw, guide the other end near the screw.
Tightly twist the seal wire
1
right behind the head hole
screw
3
to secure the screw.