CORIOLISMASTER FCX100, FCX400
SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
EN – 7
3.2
FCB400, FCH400 in integral mount design with
single-compartment housing.
NOTICE
Before the device is sealed, any settings must be
completed and the write protection switch must be set to
“write protection active” (see chapter ‘Activating the
hardware write protection’ on page 3 ).
Scope of delivery
G12355
1
Fig. 7
Item Designation
Quantity
1
Head hole screws
2
Seal 1
Seal
wire
Preparation
G12356
2x
1
1
2
Fig. 8
1. Loosen two of the four screws
2
on the foot and
replace with head hole screws
1
.
Lay seal wire
G12357
2
1
Fig. 9
1. Insert the seal wire
1
through the eyelets
2
on the
housing and cover on the left side such that both ends
are the same length.
2. Tightly twist the seal wire
1
right behind the
eyelets
2
such that the twist loops are smaller than 1
cm.
G12358
2
1
Fig. 10:
1
Seal wire
2
Eyelets
3. Continue to twist the seal wire
1
up to the right
eyelets
2
and secure as previously described.