CORIOLISMASTER FCX100, FCX400
SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
EN – 9
3.3
FCB100, FCH100, FCB400 and FCH400 sensors in
remote mount design
NOTICE
Before the device is sealed, any settings must be
completed and the write protection switch must be set to
“write protection active” (see chapter ‘Activating the
hardware write protection’ on page 3 ).
Scope of delivery
G12359
1
2
Fig. 12
Item Designation
Quantity
1
Holding
rod
1
2
Locking
cylinder
1
Seal 1
Seal
wire
Preparation
G12360
1x
2
1
3
Fig. 13
1. Unscrew the fixing screws
3
on the cover of the
housing and replace with the holding rods
2
.
Lay seal wire
G12352
2
3
1
Fig. 14
1. Insert locking cylinder
3
on the holding rod
1
on the
housing cover.
2. Insert the seal wire
2
through the openings on the
holding rod
1
such that both ends are the same length.
3. Tightly twist the seal wire
2
right behind the locking
cylinder
3
to secure the locking cylinder
3
.
NOTICE
You should still be able to slightly move the locking
cylinder after it is secured with the seal wire.
4. Both ends of the seal wire must be very close together
and the twist loops should be smaller than 1 cm.
5. Place the seal approx. 5 cm from the locking cylinder
3
onto the ends of the seal wire and press together using
sealing pliers. Cut off any excess seal wire.