CORIOLISMASTER FCX100, FCX400
SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
EN – 11
Lay seal wire
G12365
1
2
4
3
Fig. 17: Example of FCX100 and FCX400 sensor in remote mount
design
3. Continue to twist the seal wire
1
up to the rear head
hole screw
2
. Guide one end of the wire through the
hole in the screw, guide the other end near the screw.
Tightly twist the seal wire
1
right behind the head hole
screw
2
to secure the screw.
4. Continue to twist the seal wire
1
up to the front head
hole screw
3
. Secure the front head hole screw
3
as
described previously.
5. Tightly twist both ends of the seal wire after the front
head hole screw
3
over a length of approx. 5 cm.
6. Place the seal
4
approx. 5 cm from the front head hole
screw
3
onto the ends of the seal wire and press
together using sealing pliers. Cut off any excess
1
seal
wire.
3.5
FCT400 sensor with dual-compartment housing
NOTICE
Before the device is sealed, any settings must be
completed and the write protection switch must be set to
“write protection active” (see chapter ‘Activating the
hardware write protection’ on page 3 ).
Scope of delivery
G12350
1
2
3
Fig. 18
Item Designation
Quantity
1
Holding
rod
2
2
Locking
cylinder
2
3
Head hole screw
2
– Seal
1
– Seal
wire