17
Electromagnetic Flowmeter COPA-XT
5.4.
Parameter Overview and Data Entry in “Condensed Form”
Submenu/Parameter
Entry Type
Comments
from table/numeric
Data can only be entered after the program protection has been
turned off.
on/off
If a PP-Code other than “0” (factory setting) has been stored, then
the program protection can only turned off after the appropriate
PP-Code (1-255) has been entered.
When the
program protection is turned off, parameters can be
changed.
numeric
When the
program protection is turned off, it is possible to enter a
new PP-Code.
Enter old PP-Code
0 = factory setting
Enter new PP-Code (0-255)
from table
English, German
The meter size of the flowmeter
See Instrument Tag
Automatic flow range end value setting determined form the meter
size selection
Flow range end value can be set from
0.05 Q
max DN
to 1 Q
max DN
Flow range for forward and reverse flow directions
Min. flow range setting 0 - 0.5 m/s
Max. flow range setting 0 - 10 m/s
Flow range end value can be set between 0.5 and 10 m/s
For int. and ext. flow totalization, range 0.001 - 1000 pulses
per selected units, max. count Frequency 100 Hz.
numeric
For external pulse output,
Range 0.1 ms - 2000 ms
* Prog. Protection *
on
ENTER
* Prog. Protection *
off
PP-Code?
0
* Prog. Protection *
off
Prog. Prot. Code
ENTER
Old PP-Code?
0
New PP-Code?
0
Language
English
Meter Size
DN 50
2 In
Q
max DN
10 m/s
60.00
m3/h
Qmax
12.000
m3/h
Pulse
1.0000
/m3
Pulse Width
30.000 ms