Electromagnetic Flowmeter COPA-XT
11
3.3.3 Interconnection Diagram COPA-XT
Functional ground
C9
E9
61
G2
V9
V8
TW+
TW-
a)
b)
1)
2)
1) Supply power from transmitter power supply, TW+, TW-, 14-55 V DC
2a) Contact output, function configurable in the software for system monitoring,
or F/R flow direction signal*). Output closed at alarm. Function C9, E9 or
b) Scaled pulse output, passive, pulse width can be set from 0.1 ms to 2000 ms,
terminals: V8, V9 Function 61, G2, optocoupler, fmax 100 Hz.
Optocoupler, passive, terminals V8, V9
0 V
≤
U
CEL
≤
2 V, 16 V
≤
U
CEH
≤
30 V
0 mA
≤
I
CEH
≤
0.2 mA, 2 mA
≤
I
CEL
≤
220 mA
*) The factory setting is no function for binary contact output.
Comment: To maintain the EMC-Requirements the instrument must be connected to ground.
Fig. 24
Interconnection Diagram COPA-XT