6.0 MAINTENANCE
6.1 General
Except for an occasional performance verification check, there is no routine maintenance required for
the meter. It is suggested that the meter body not be disassembled. If disassembled, complete
waterproof sealing is required for satisfactory operation and is best done at the factory. Replacement
of faulty magnet drive coils and electrode replacement is a factory operation. Factory calibration after
this type of repair is the only way to guarantee meter accuracy.
The manufacturer offers a Repair/Exchange Program to facilitate replacement of a defective meter or
converter. If the equipment is beyond the warranty limit, under this program a fixed price will be
charged for replacement of defective equipment, with appropriate credit issued when the repairable
unit is received by the manufacturer (charges prepaid). The equipment available under this program
is as follows:
•
the meter with the interconnection junction box without the signal converter
•
the signal converter
WARNING
All flowmeters and/or signal converters being returned to the
manufacturer for repair must be free of any hazardous materials (acids,
alkalis, solvents, etc.). A Material Safety Data Sheet (MSDS) for all
process liquids must accompany returned equipment. Contact the
manufacturer for authorization prior to returning equipment.
NOTE
Operation and maintenance procedures for the signal converter are pro-
vided in the instruction bulletin supplied with the signal converter.
When communicating with the manufacturer in regard to replacement of a complete meter or the
signal converter, it is important to refer to the complete instrument serial number to assure that the
correct replacement will be supplied. This information is provided on the manufacturing specification
sheet supplied with the meter, and on the instrument data tag.
6.2 System Troubleshooting
In the event that faulty operation of the meter is evident, the following procedure can be used as a
guide to isolate the malfunctioning device as either the meter or the signal converter. A standard
multimeter is suitable for making most of the test measurements.
To supplement the following discussion refer to:
•
Section 5.0 Circuit Description
•
Signal converter instruction manual
10DS3111E INSTRUCTION MANUAL