CI/ACX-EN Rev. D
ACX Analyzer System Commissioning Instruction
17
Analyzer System Installation Site Requirements,
continued
Dimensions and Space
Requirement
Refer to the “Layout Plan” in the drawings set.
Installation Site
Stability
The installation site floor must be plane and capable of supporting the cabinets
weight (see page 22).
The installation site wall must be capable of supporting the weight of the mounting
plate and the separate electrical distribution cabinet (see page 22).
Sample Gas Inlet Conditions
(at the Extraction Point)
CAUTION!
The analyzer system must not be used for measurement of flammable gases and
ignitable gas/air or gas/oxygen mixtures!
In the case of toxic gases, the threshold limit value (TLV) must be complied with.
Application
Temperature
Pressure p
abs
Flow
Emission Monitoring
max. 500 °C
850 to 1100 hPa (0.85 to 1.1 bar)
max. 250 l/h
Kiln or Calciner Outlet (T > 900 °C)
max. 1300 °C
850 to 1100 hPa (0.85 to 1.1 bar)
1)
max. 100 l/h
Calciner
max. 900 °C
850 to 1100 hPa (0.85 to 1.1 bar)
max. 125 l/h
Wet Kiln Gas Outlet
max. 300 °C
850 to 1100 hPa (0.85 to 1.1 bar)
max. 250 l/h
2)
Preheater/CO Monitoring of ESP
max. 450 °C
850 to 1100 hPa (0.85 to 1.1 bar)
max. 250 l/h
3)
Coal Bunker, Coal Mill
max. 500 °C
850 to 1100 hPa (0.85 to 1.1 bar)
max. 250 l/h
Process Measurement
max. 500 °C
850 to 1100 hPa (0.85 to 1.1 bar)
max. 250 l/h
1) at the sample gas inlet of the analyzer system
2) max. 60 l/h with SO
2
measurement
3) max. 300 l/h with probe F
Test Gas Connection at the Gas Sampling Probe or upstream of the
Sample Gas Cooler: Test Gas Inlet Conditions
CAUTION!
When handling test gases, the lower explosion limit (LEL)as well as the threshold
limit value (TLV) must be complied with.
Specification
Pressure p
e
Flow
Test gases 1, 2, 3
Sample component or substitute gas
component in N
2
or air
1000
±
100 hPa
(1.0
±
0.1 bar)
130 to 250 l/h