ACX Analyzer System Commissioning Instruction
CI/ACX-EN Rev. D
Program Sequence
PFE2
Digital output:
-D08
DO2
-D08
DO3
-D08
DO1
-D08
DO4
-D08
DO5
-E05
MV1
Display
Status
signal
Valve:
-Y1.1
-Y2.1
-Y3.1
-Y4
-Y5
6)
-Y01
Message
“Purge
back
active“
Maint.
Mode
Step Duration Function
Impulse
Compr.
Air
Filter
Back-
purging
Tube
Back-
purging Venting
Sample
Gas
Valve
Position
Calibr.
Valve
5)
0
4 hrs
1)
Measure
closed
closed
closed
open
open
Measure off
off
1
10 sec
8)
Back-purging
probe filter
Impulse open
closed
closed
closed
Calibrate on
on
2
14 sec
9)
Back-purging
probe tube
Impulse closed
open
closed
closed
Calibrate on
on
3
6 sec
Venting
closed
open
closed
open
closed
Calibrate on
on
4
150 sec
2)
Post-purging closed
closed
closed
open
open
Measure on
on
0
4 hrs
1)
Measure
closed
closed
closed
open
open
Measure off
off
PFE3, Probe 2,
Probe F, Probe
Tube 40W
Digital output:
-A01
DO1
-A01
DO2
-A01
DO3
-E05
MV1
Display
Status
signal
Valve:
-Y12
7)
-Y11
-Y07
-Y01
Message
“Purge
back
active“
Maint.
Mode
Step Duration Function
Filter
Back-
purging
Tube
Back-
purging Venting
Position
Calibr.
Valve
5)
0
4 hrs
1)
Measure
closed
closed
closed
Measure off
off
1
2 sec
Switch over
closed
closed
closed
Calibrate on
on
2
4 sec
3)
Back-purging
probe filter
open
closed
closed
Calibrate on
on
3
8 sec
4)
Back-purging
probe tube
closed
open
closed
Calibrate on
on
4
6 sec
Venting
closed
closed
open
Calibrate on
on
5
150 sec
2)
Post-purging
closed
closed
closed
Measure on
on
0
4 hrs
1)
Measure
closed
closed
closed
Measure off
off
1) Cycle time factory-set to 4 hours
2) Post-purging time factory-set to 150 sec. For an analyzer system with Probe F,
this period must be as short as possible (to be determined during start-up)
3) 1x pressure impulse 2 sec, 1x interrupt 2 sec
4) 1x pressure impulse 2 sec, 1x interrupt 2 sec, 1x pressure impulse 4 sec
5) Calibration valve on = “Measure”, calibration valve off = “Calibrate”
6) Only in version with VOC analyzer AO2000-Fidas24:
open = sample gas path open, closed = relaxation against atmosphere
7) not in a system with Probe 2, Probe F
8) 3x interrupt 2 sec, 2x pressure impulse 2 sec
9) 3x interrupt 2 sec, 2x pressure impulse 2 sec, 1x pressure impulse 4 sec
Continued on next page