ACX Analyzer System Commissioning Instruction
CI/ACX-EN Rev. D
Adjustment of Cycle Time and Post-Purge Time
Adjustment of Cycle
Time and Post-Purge
Time
To adjust the cycle time or post-purge time you must change the parameters of
function blocks.
WARNING!
Only changes as described below may executed! Changes of function block
parameters inappropriately executed may affect the complete function of the
function blocks program!
Procedure
Step
Action
1
Push Softkey
MENU
.
The window
MAIN MENU
is shown.
2
Select menu
Configure
.
The window
CONFIG:
is displayed.
3
Select menu
Functions blocks
.
The window
CONFIG: FUNCTION BLOCK
is displayed.
4
Select menu
Miscellaneous
and after this menu
Timer
.
The window
CONFIG.: TIMER
is displayed.
To adjust the cycle time:
5
Select timer
Zycl.
.
The window
CONFIG: TIMER CYCL.
is shown with the parameters
of this function block.
6
Select the parameter
Low time
.
The window
PASSWORD ENTRY
is shown, if the password is not
already active.
7
Enter the password, using the numeric keys. The factory-set password is
325465.
The window
CONFIG: TIMER CYCL.
with indication of the
parameter
Cycle time
is shown.
8
Change the shown value (factory-set = 4 hrs) to the new value required.
9
Return to normal measuring operation using the key
MEAS
.
To adjust the post-purge time:
5
Select timer
DELAY
.
The window
CONFIG: TIMER DELAY
is shown with the parameters
of this function block.
6
Select the parameter
Low time
.
The window
PASSWORD ENTRY
is shown, if the password is not
already active.
7
Enter the password, using the numeric keys. The factory-set password is
325465.
The window
CONFIG: TIMER DELAY
with indication of the
parameter
Low time
is shown.
8
Change the shown value (factory-set = 150 sec) to the new value required.
9
Return to normal measuring operation using the key
MEAS
.