14
Incorrect code.
or
SET UP INPUT
Security Code -----
Nitrate
<Units>
CALIBRATION PASSED
or
CAL FAILED (SLOPE)
or
CAL FAILED (SLOW)
or
Mode
Calibrating Std 2
Electrode mV
– –
•
–
Cal
Nitrate <units>
…6
PROGRAMMING
…6.2.2
Operating Page 2
Continued from previous page.
Calibrating Standard 2
The upper display shows the ammonia concentration value. The display remains
until a stable output is obtained from the probe.
Toggle between the two displays.
Electrode Millivolts
The probe output can also be displayed during calibration
Note.
Pressing
Cal
during a calibration aborts the sequence and returns to
normal operation.
When a stable probe output is detected the display automatically advances to the
next parameter.
Calibration Passed
A satisfactory calibration has been carried out.
or
Calibration Failed (slope)
The monitor was unable to obtain an adequate slope value during calibration. The
slope value required is 80 to 110%.
or
Calibration Failed (slow)
The monitor was unable to obtain a stable output from the probe during calibration.
Return to top of
Operating Page 2.
or
Return to
Operating Page 1.
6.2.3
Security Code Page
A security code is required to gain access to all subsequent programming pages. The code is preset at the factory to '0' but can
be changed if required in the Current Output Page – see Section 6.2.5.
Security Code
Set the correct security code.
Advance to the first of the programming pages –
Set Up Input Page.
Return to
Operating Page 1.