8 Error messages
LCD Display messages
The LCD HMI in case of transmitter errors or malfunctioning is capable of displaying specific error/fault messages to help the user
in identifying the problem and resolve it. In case of an alarm, a message consisting of an icon and text appears at the bottom of the
process display, as shown hereafter. Use the (1) key to call up the information level. Use the “Diagnostics” menu to call up the error
description with a help text. In the error description, the error number is displayed in the second line (M028.018). Two further lines
are used to describe the error. The device status is divided into four groups. The message text beside this icon in the display
provides information about where to look for the error. There are the following areas: Electronic, Sensor, Configuration,
Operanting and Process.
Icon
Description
Error / Failure
Functional check (e.g. during simulation )
Out of Spec (e.g. operating with empty meter pipe)
Maintenance required
Error states and alarms
Error code
Description
Possible cause
Suggested /recommended action
Current output
F114.000
The sensor signal value of the
Primary unit is incorrect due to a
mechanical failure.
Mechanical damage to the sensor.
Loss of fill fluid from the cell,
ruptured diaphragm, broken
sensor, etc
The Primary unit sensor must be replaced.
UP/DW ALARM
F110.001
The measurement accuracy could
be decreased more than an
acceptable error.
The circuitry for the sampling of the
Primary unit temperature has failed.
The Primary unit sensor must be replaced.
UP/DW ALARM
F113.002
The sensor signal value of the
Secondary unit is incorrect due to a
mechanical failure.
Mechanical damage to the sensor.
Loss of fill fluid from the cell,
ruptured diaphragm, broken
sensor, etc
The Secondary unit sensor must be replaced.
UP/DW ALARM
F109.003
The measurement accuracy could
be decreased more than an
acceptable error.
The circuitry for the sampling of the
Secondary unit temperature has
failed.
The Secondary unit sensor must be replaced.
UP/DW ALARM
F102.004
The Process value of the Primary
unit is outside the sensor limits and
no longer representing the true
applied process value.
The measurement range has not
been correctly calculated OR an
incorrect transducer model has
been selected.
The compatibility of pressure transmitter model
and process conditions has to be checked.
Probably a different transmitter type is required.
no effect
S054.005
The sensor temperature of the
Primary unit is outside of its
operational limits.
The temperature of the process
environment affects the pressure
transmitter. Excess temperature
can reduce accuracy, degrade
device components and may require
calibration/replacement.
The compatibility of pressure transmitter model
and process conditions has to be checked. A
different installation type could be required e.g.
use of remote seals.
no effect
F101.006
The Process value of the Secondary
unit is outside the sensor limits and
no longer representing the true
applied process value.
The measurement range has not
been correctly calculated OR an
incorrect transducer model has
been selected.
The compatibility of pressure transmitter model
and process conditions has to be checked.
Probably a different transmitter type is required.
no effect
S053.007
The sensor temperature of the
Secondary unit is outside of its
operational limits.
The temperature of the process
environment affects the pressure
transmitter. Excess temperature
can reduce accuracy, degrade
device components and may require
calibration/replacement.
The compatibility of pressure transmitter model
and process conditions has to be checked. A
different installation type could be required e.g.
use of remote seals.
no effect
Summary of Contents for 266 DDS
Page 30: ...Device Set up 30 266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B ...
Page 31: ...266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B 31 ...
Page 32: ...32 266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B ...
Page 33: ...266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B 33 ...
Page 34: ...Display 34 266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B ...
Page 35: ...266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B 35 ...
Page 36: ...36 266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B ...
Page 41: ...Device Info 266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B 41 ...
Page 42: ...Communication 42 266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B ...
Page 59: ...266 DDS HART PRESSURE TR ANSMITTERS OI 266DDS EN REV B 59 Notes ...