VETSCAN SA USER’S MANUAL
v
1.1 SOFTWARE
|
3 4
Clean Fan Filter
Check the Fan Filter on the Side Panel Side Door periodically to ensure proper ventilation of the
SA analyzer. Check the Fan Filter frequently if the analyzer is in an environment with excessive
dust or animal hair.
1. Remove the Fan Filter (do not run sample measurement during this cleaning).
2. Rinse Fan Filter with warm water to remove debris. Do not remove the foam from the
plastic frame.
3. Dry the Fan Filter to remove excess water and place it back into the analyzer.
For replacement Fan Filter contact your local Abaxis representative.