![background image](http://html1.mh-extra.com/html/abaxis/vetscan-sa/vetscan-sa_user-manual_450825021.webp)
VETSCAN SA USER’S MANUAL
v
1.1 SOFTWARE
|
2 1
WARNING!
If the Strip Tray has already moved into the analyzer to start scanning before the
strip is placed in the tray, an error will be reported on the UA and sent to the SA. For best
results, a new strip will have to be wetted to restart the testing workflow.
While measuring, the SA screen is locked and navigation to other screens is prohibited. Also,
the UA status indicator light (top right of the UA Results Panel) will display yellow during
measurement process. Do not disconnect UA from SA during testing process.
The UA analyzer can operate independently from the SA. However, this is not advised if combined
urinalysis reports are desired or if UA test results are needed to be stored in the SA database.
Operating the UA independently from the SA indicates that the test was not initiated from the
SA or that the UA was disconnected from the SA. If UA test results need to be transferred to the
SA, the UA will need to be connected to the SA. Next, navigate to the UA Database and select
the desired test report to transmit and push the “USB send” icon to transmit the data to the SA.
The UA test results can be combined with SA test results in Database menu.