VETSCAN SA USER’S MANUAL
v
1.1 SOFTWARE
|
2 6
Retest SA
If one of these 3 SA Warnings appear, a restest mode is enabled:
1. Review images
2. TNTC – Dilute and retest
3. Possible short sample
The SA analyzer provides an immediate option to retest the sample without any additional
patient entry. The retest function is only available for the SA test. If a UA test was also
completed during the original test run, the UA results will be saved and displayed with the new
SA test values.
SA Results displayed on Measure screen after test completed.
If a retest on the SA is not desired, just click “Save”. If a retest is desired, the initial SA test data
will not be saved to the Database. Follow these steps to retest:
1. Click Retest button; all the inputs to the sample will automatically be populated.
2. Select Dilution if necessary:
Dilute the sample and select correct dilution. Undiluted is the default setting and
appropriate for most samples. Diluted samples are to be used on the SA analyzer only.
Verify that all patient and sample information is correct.