![background image](http://html.mh-extra.com/html/aandd/winct-plus/winct-plus_instruction-manual_446567006.webp)
5
4. Network Setting
To implement a LAN connection, it is required to set the IP address, subnet mask, etc. for both
the computer and the OP-08. Please consult with the network administrator before assigning IP
addresses, etc.
OP-08 Factory setting
IP address
172.16.100.2
Subnet mask
255.255.0.0
Port number
10001
Attention
-Do not connect the OP-08 to a backbone network.
If any failure in network should occur due to the OP-08, A&D Co., Ltd. will not assume any
responsibility.
-Before connecting to an existing network, consult with the network administrator.
Setting example
Example 1) Connecting two GH balances.
For LAN cabling, straight cables should be used.
IP address
Subnet mask
Computer
172.16.100.1
GH-252
172.16.100.2
GH-300
172.16.100.3
255.255.0.0
Example 2) Connecting a computer directly to a single GH balance
For LAN cabling, cross cables should be used.
IP address
Subnet mask
Computer
172.16.100.1
GH-252
172.16.100.2
255.255.0.0
Ethernet
hub
Computer
172.16.100.1
172.16.100.2 172.16.100.3
Straight cables
172.16.100.1 172.16.100.2
Computer
Cross cable