13
The Property window opens.
Follow the steps below to set up:
①
Provide a name for the connection, e.g. GH-252 (the name of the weighing instrument).
②
Check the LAN option to select LAN connection.
③
Enter the IP address of the OP-08.
④
Enter the port number of the OP-08. In the default setting, it is set at 10001.
⑤
Fill in the “Send Data Command” and “Terminator” boxes to send to the weighing instrument.
-For collecting data by sending a send data command to the weighing instrument.
Enter the “Send Data Command” box.
-For collecting data by pressing the [PRINT] key on the weighing instrument.
(When the weighing instrument is used in Key mode, Auto print mode and Stream mode.)
Blank out the “Command” box and enter the “Terminator” box only.
⑥
Click the “Test” button to check that the communication is made.
If the “Command” box is filled in the step 5 above, a reply to the command will be displayed.
If it is blank, send data by pressing the [PRINT] key on the weighing instrument to check the
display.
If the data is displayed as shown in
⑧
above, the communication is normally made.
If no data is displayed, check the settings again.
⑦
Click the OK button to complete the setting.
①
②
③
④
⑤
⑥
⑦
⑧