SV-10A/SV-10H output format example
Viscosity
/
Temper-
ature
Display
Output format example
Remarks
L
LLLL
mPa
・
s
+00000.00,mPa s,+025.67,C
Zeroes are output for below measuring
range error.
0.30 mPa
・
s
+00000.30,mPa s,+025.67,C
10.0 mPa
・
s
+00010.00,mPa s,+025.67,C
100 mPa
・
s
+00100.00,mPa s,+025.67,C
1.00 Pa
・
s
+01000.00,mPa s,+025.67,C
For 1000 mPa
⋅
s or greater, the displayed unit
is Pa
⋅
s, but the output unit remains mPa
⋅
s.
M
mPa
・
s
/
°C
H Pa
・
s
+12000.00,mPa s,+025.67,C
12000 is output for above measuring
range error.
L Pa
・
s
+000.0000, Pa s,+051.23,F
Zeroes are output for below measuring
range error.
0.0003 Pa
・
s
+000.0003, Pa s,+051.23,F
0.0100 Pa
・
s
+000.0100, Pa s,+051.23,F
0.100 Pa
・
s
+000.1000, Pa s,+051.23,F
1.00 Pa
・
s
+001.0000, Pa s,+051.23,F
Pa
・
s
/
°F
H Pa
・
s
+012.0000, Pa
,+051.23,F
12 is output for above measuring range error.
L cP
+000.0000, cP
,+025.67,C
Zeroes are output for below measuring
range error.
0.30 cP
+00000.30, cP
,+025.67,C
10.0 cP
+00010.00, cP
,+025.67,C
100
cP
+00100.00, cP
,+025.67,C
10.0 P
+01000.00, cP
,+025.67,C
For 1000 cP or greater, the displayed
unit is P, but the output unit remains cP.
C
cP
/
°C
H P
+12000.00, cP
,+025.67,C
12000 is output for above measuring
range error.
L
P
+000.0000,
P
,+051.23,F
Zeroes are output for below measuring
range error.
0.0030 P
+000.0030,
P
,+051.23,F
0.100 P
+000.1000,
P
,+051.23,F
1.00 P
+001.0000,
P
,+051.23,F
10.0 P
+010.0000,
P
,+051.23,F
P
/
°F
H P
+120.0000,
P
,+051.23,F
120 is output for above measuring range error.
: Space (ASC 20h)
SV-100A/SV-100H output format example
Viscosity
/
Temper-
ature
Display
Output format example
Remarks
L Pa
・
s
+0000.000, Pa s,+025.67,C
Zeroes are output for below measuring
range error.
1.00 Pa
・
s
+00001.00, Pa s,+025.67,C
10.0 Pa
・
s
+00010.00, Pa s,+025.67,C
Pa
・
s
/
°C
H Pa
・
s
+00120.00, Pa
,+025.67,C
120 is output for above measuring range
error.
L P
+000000.0,
P
,+051.23,F
Zeroes are output for below measuring
range error.
10.0 P
+000010.0,
P
,+051.23,F
100 P
+000100.0,
P
,+051.23,F
P
/
°F
H P
+001200.0,
P
,+051.23,F
1200 is output for above measuring range error.
: Space (ASC 20h)