48
8-2-3 Simplified Calibration Using Purified Water (SV-1A/1H/10A/10H)
1 Place the purified water in the sample cup.
2 Press the START key to measure
the purified water. Confirm that the
viscosity and temperature values are
stabilized.
3 Press and hold the START key.
The theoretical viscosity value
(Viscosity
×
Density)
of the purified
water at the measuring temperature
is displayed and all the displays
blink.
To cancel the operation, press the
STOP key. The display returns to
the status before calibration.
4 Press the START key again to
perform calibration.
When calibration is completed, "
end
"
is displayed. Then, the viscosity
returns to the measuring status.