9-3 Description of Items (For all models)
Condition (
Cond
)
The stability of the viscosity measurement results can be adjusted, taking ambient conditions such
as vibration into consideration.
Parameter
Settings Description
0
Follows the viscosity changes
quickly.
(Prone to vibration)
1
•
2
Follows the viscosity changes
slowly.
(Stable values)
When the viscosity value is unstable due to external
vibration, set a greater parameter.
To measure while following the rapid changes in
viscosity, set a smaller parameter.
With a smaller setting, the measurement is prone to
external vibration. Consider the ambient conditions
of the installation site.
Unit Upon Power-on (
Unit
)
The units of viscosity and temperature displayed when the power is turned on are specified.
SV-1A/1H/10A/10H
Parameter
Settings Description
0
•
m
mPa
⋅
s
(Millipascal second)
1
Pa
⋅
s
(Pascal second)
2
c
cP
(Centipoise)
3
P
(Poise)
°C
(Celsius)
4
mPa
⋅
s
(Millipascal second)
5
Pa
⋅
s
(Pascal second)
6
c
cP
(Centipoise)
7
Viscos-
ity
P
(Poise)
Temper-
ature
°F
(Fahrenheit)
In the standby mode, pressing
the MODE key switches the
viscosity unit.
mPa
⋅
s
⇔
Pa
⋅
s, cP
⇔
P
With "
fnc 0
" selected, units can
be switched even during
measurement. *1
∗
1 While receiving the measurement data using the graphing program RsVisco, changing the
mode using the HOLD key is not available.
For the viscosity over 1000mPa
⋅
s, the unit is fixed to Pa
⋅
s. For the viscosity over 1000cP, the unit
is fixed to P.
Decimal Point (
pnt
)
Parameter
Settings Description
0
•
Dot
"
.
"
1
Comma "
,
"
The decimal point format for the displayed measurement data and the
decimal point code for measurement data output via RS-232C are
specified.
With "Comma" selected, the separator for CSV format and RsVisco
format will be "
;
" (semicolon).