33
6-3
Basic Measurement Procedure (For the SV-H series)
1 Pour the sample into the beaker (500ml) etc.
2 With the power turned off, press the ON:OFF key.
When the display is in the standby mode, the SV-H
can measure.
3 Hold the SV-H handle firmly in your hand. Prepare to
insert the sensor plates in the sample.
4 Press the START key. Insert the sensor plates into the sample.
The measurement time is about 15 seconds. After inserting the sensor plates in the sample,
maintain the height of the sensor plates.
l
is displayed when the sensor plates do not touch the liquid.
Right after the sensor plates touch the liquid,
l
will be still displayed. Continue the
measurement with this state. After stabilizing, the viscometer displays the viscosity value.
[Precautions when inserting the sensor plates into the sample]
Confirm that the protector is in the position as shown in the figure, and then insert the sensor
plates into the sample.
Confirm that the sample surface and the center of the narrow part of the sensor plates match up.
Keep the SV-H at this position (Height).
At this time
- With SV-1H, there are the triangular marks on the sensor plate. Confirm that the sample
surface is between the upper and lower triangular marks.
Beaker
Sample
(Standby mode)
Protector
Sensor plates
Summary of Contents for SV-1A
Page 6: ...4...