6 Confirm that the positioning of the sensor plates is in center of
the sample cup by adjusting the X-Y-Z stage.
7 Turn the knob (Height direction) so
as to adjust the sample surface to
the center of the narrow part of the
sensor plates. At this time, use the
surface locator plate as a guide. The
surface locator plate has been
secured in position so that the tip of
the surface locator plate comes into
contact with the sample surface.
Note
•
Adjust the height of the sample surface correctly, otherwise it may cause a
measurement error.
•
With SV-10A/SV-100A, the surface locator plate can be attached or removed by loosening the
screw.
•
With SV-10A/SV-100A, before removing the sensor protective cover, remove the
surface locator plate.
•
Removing and reattaching the surface locator plate will cause the position (Height) of
the sensor plates and the sample surface to change. Therefore, it is recommended
that calibration be performed using a standard viscosity fluid before measurement.
Note
Use the protector in the position as shown on the left below. If the protector is not used, a
measurement error may occur, especially in measuring a viscosity over 5000 mPa
⋅
s.
When the position of the sensor plates in the liquid is not at the same level, level the
viscometer using the leveling feet so that the liquid surface will be leveled.
Sensor protective
cover
Surface
locator plate
(Shaded portion)
Knob
(Height direction)
Sample
surface
adjustment
position