![A&D MC Series Instruction Manual Download Page 61](http://html2.mh-extra.com/html/aandd/mc-series/mc-series_instruction-manual_3084477061.webp)
59
17. SPECIFICATIONS
MC-1000
MC-6100
Weighing capacity
1100 g
6100 g
Maximum display
1100.0844 g
6100.844 g
Minimum weighing value
0.0001 g
0.001 g
Repeatability (Standard deviation)
See Note 1 below.
0.0005 g / 1 kg
0.0004 g / 500 g
0.004 g / 5 kg
0.0015 g / 2 kg
Sensitivity drift
(10°C-30°C/50
°
F-86
°
F with )
±2 ppm/°C (When automatic self calibration is not used)
±
0.0100 g
±
0.150 g
Accuracy right after calibration using
the internal mass. See Note 2 below
The value above is to the weighing capacity.
Operating environment
5°C to 40°C (41°F to 104°F)
85%RH or less (No condensation)
Internal mass
Yes
Data memory function, weight data
Maximum 200
(Maximum 100 when the time and date are added)
Calendar function
Yes
Display refresh rate
5 times/second or 10 times/second
Display mode
g (gram)
Interface (Provided as standard)
RS-232C with Windows Communication Tools WinCT
External calibration weight
1000 g, 900 g
800 g, 700 g
600 g, 500 g
400 g, 300 g
200 g
6000 g
5000 g
4000 g
3000 g
2000 g
Weighing pan
128 x 128 mm
165 x 165 mm
Net weight
Approx. 4.6 kg
Approx. 5.1 kg
External dimensions
210 (W) x 317 (D) x 86 (H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage
and power receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Accuracy class F1
500 g, 1 kg
2 kg, 5 kg
Note 1: When the auto-centering pan or automatic system is used in good ambient conditions.
Note 2: Accuracy right after calibration using the internal mass in good ambient conditions (within
the temperature range of 10
°
C to 30
°
C (50
°
F to 86
°
F) with no abrupt changes in
temperature or humidity, no drafts, no effect by magnetic fields or static electricity).
The value of the internal mass may change due to corrosion or other damage caused by the
operating environment, or due to aging. Check the internal mass using an external weight
periodically.