![A&D MC Series Instruction Manual Download Page 39](http://html2.mh-extra.com/html/aandd/mc-series/mc-series_instruction-manual_3084477039.webp)
37
CSV format
5if type 5
z
Separates the data of A&D standard format and the unit by a comma (,).
z
Outputs the unit even when the data is overloaded.
z
When the ID number, data number, time and date are added, outputs the ID number, data
number, date, time and weighing data in this order and separates each item by a comma and
treats all the items as one group of data.
Note
To add the ID number, data number, time and date, the function settings must be changed.
9-7 Description of the Data Format Added to the Weighing Data
ID number
dout 5-id 1
The number to identify a specific balance.
z
This format consists of seven characters excluding the terminator.
Data number
dout d-no 1
This format outputs the data number just before the data is transmitted using the RS-232C interface.
z
This format consists of six characters excluding the terminator.
z
When CSV format (
5if type 5
) is selected, the period (
.
) is replaced with a comma (
,
).
Note
The data number is added only to the weighing data that is stored in memory.
Date
dout 5-td 2
or
3
z
The date output order can be changed in “Clock (
Cl adj
)”. Outputs the year in four-digit
format.
Time
dout 5-td 1
or
3
z
Outputs time in 24-hour format.
Note
When the data described above is added to the weighing data, the output is in the following
order: ID number, Data number, Date, Time and Weighing data.
2 0 0 9 / 1 2 / 3 1
CR LF
CR LF
O L , +
S T , + 1 0 0 0 . 0 0 0 0 ,
g
CR LF
LAB-123, No,012, 2009/12/31, 12:34:56, ST,+1000.0000, g<CR><LF>
ID number Data number Date
Time
Weighing data
9 9 9 9 9 9 9 E + 1 9 ,
g