24
7-6 Correcting the Internal Mass Value
The MC series balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected mass value
is maintained in non-volatile memory even if the AC adapter is removed.
The internal mass value is corrected as follows:
Model
Correction reference value
Correction range
MC-1000
500.0000 g
±0.020 g
MC-6100
2000.000 g
±0.20 g
Example: Using the MC-6100
Operation
1 Calibrate the balance using the internal mass (one-touch
calibration). Then, place an external weight and confirm
the value to be corrected.
In the example, the value is off by -0.060 g and the
correction reference value of the MC-6100 is 2 kg. So, the
value is to be corrected by 0.030 g in 2 kg.
2 Press the ON:OFF key to turn off the display.
3 While pressing and holding the PRINT key and the
SAMPLE key, press the ON:OFF key. The balance
displays
p5
.
Correct the internal mass
by +0.030 g in 2 kg.
Calibrate the balance with
the corrected internal mass.
4 kg
4 kg
Use the same weight
With these keys
held down
Press the key