![A&D MC Series Instruction Manual Download Page 37](http://html2.mh-extra.com/html/aandd/mc-series/mc-series_instruction-manual_3084477037.webp)
35
Interval memory mode
The weighing data is periodically stored in memory.
Required setting
dout prt
3
Interval memory mode
dout data
2
Data memory function is used.
Stores weighing data.
dout int
Interval
time
Optional setting
dout
5-td1
,
2
,
or
3
Adds the time and date.
Example
“For periodical weighing without a computer command and
outputting all of the data to a computer at one time”
9-6 Description of the Item “Data format”
A&D standard format
5if type 0
This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is
used, set the printer to MODE 1 or 2.
z
This format consists of fifteen or sixteen characters excluding the terminator.
With the MC-1000, when the numerical data excluding the decimal point exceeds eight
characters, the data format will be sixteen characters long.
z
A header of two characters indicates the balance condition.
z
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is used.
z
The unit, consisting of three characters, follows the data.
(Data example exceeding 8 characters)
DP (Dump print) format
5if type 1
This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B
is used, set the printer to MODE 3.
z
This format consists of sixteen characters excluding the terminator.
z
A header of two characters indicates the balance condition. No overload header is used.
z
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
z
The unit, consisting of three characters, follows the data.
W T + 1 0 0 0 . 0 0 0 0
Terminator
CR LF
Header
Data
Unit
Unit
Data
Header
S T , + 0 0 0 . 0 1 2 7
8 8
g
CR LF
Terminator
Unit
Data
Header
S T , + 1 0 0 0 . 0 0 0 0
8 8
g
CR LF
Terminator