9
Auto-centering pan (AX-MC6100PAN ) installation procedure
Caution: During installation, do not apply excessive force to the balance.
z
When used as a mass comparator, the balance can achieve even more precise weighing,
by using the auto-centering pan (sold separately), which reduces eccentric loading errors.
1. At step 2 of the glass breeze break installation procedure, secure the breeze break provided with
the AX-MC6100PAN option using the breeze break securing screws provided with the option. At
this time make sure that the side of the breeze break with the screws is placed at the back.
Attach the pan support receptors to the shock absorbers.
2. Place the auto-centering pan guide on the pan support receptors, and then place the
auto-centering pan on the guide.
Note
When replacing the weighing pan with
the auto-centering pan, be sure to
calibrate the balance before weighing.
Refer to “7. CALIBRATION”.
Breeze break
securing screw
(2 pieces)
Auto-centering pan
Option
AX-MC6100PAN
breeze break
Pan support
receptors
(4 pieces)
Auto-centering pan guide
When placing on the pan
support receptors, tilt the
auto-centering pan guide
to the left and right,
forward and backward.
Side with screws