Page 26
HV-G/HW-G Series
7. Basic Operation
7.3. Weighing Range for the HV-G Series
This is the function to select a weighing range for the HV-G series.
The mass value is displayed within a selected range.
Select automatic range (
f2 0
) or manual range (
f2 1
) in the function table.
Operation and Performance
Function table
Description
A
f2 0
Automatic range
The weighing range changes automatically, if the weight value
proceeds from narrow range to wide range when articles are
placed on the pan.
When there is nothing on the pan and the zero point mark is
displayed, the weighing range changes to the minimum range
automatically.
When the zero point mark is not displayed due to net display
or zero error, press the
ZERO
key, with nothing on the pan, to
change to the minimum range.
A
A
A
A
A
A
A
A
A
f2 1
Manual range
Press
the
RANGE
key to expand the range.
Press
the
RANGE
key to change to the minimum range, when
there is nothing on the pan and the zero point mark is
displayed.
When the zero point mark is not displayed due to net display
or zero error, press the
ZERO
key and the
RANGE
key, with
nothing on the pan, to change to the minimum range.
A
A
A
A
A
A
A
A
Weighing Range
Models Weighing
range
A
HV-15KGV, HV-15KGL
3 kg,
6 kg ,
15 kg
A
HV-60KGV, HV-60KGL
15 kg,
30 kg,
60 kg
A
HV-200KGV, HV-200KGL
60 kg,
150 kg,
220 kg
A
Summary of Contents for HV-15KGL
Page 19: ...HV G HW G Series Page 17 6 Description of Each Part 6 Description of Each Part ...
Page 20: ...Page 18 HV G HW G Series 6 Description of Each Part ...
Page 67: ...HV G HW G Series Page 65 17 Options For models with weighing capacity of 300 kg to 1200 kg ...
Page 81: ...HV G HW G Series Page 79 18 Specifications Dimensions Unit mm ...
Page 82: ...Page 80 HV G HW G Series 18 Specifications Unit mm ...
Page 84: ...Page 82 HV G HW G Series MEMO ...
Page 85: ...HV G HW G Series Page 83 MEMO ...
Page 86: ...Page 84 HV G HW G Series MEMO ...